Title:Analysis of familiar defects of low-temperature forging on operating instruments
Authors: DONG Jia-kun YE Xun SUN Bin(Operating Instrument Factory Shanghai Medical Instrument Group Co.Ltd. Shanghai 200072 China) 
Unit:  
KeyWords: operating instrument low-temperature forging forging defect 
ClassificationCode:TG316
year,vol(issue):pagenumber:2005,30(6):15-17
Abstract:
Forging defects of mould cavity non-saturate,crack,pucker and shearing edge avulsion,etc.on the astrictive clamps and operating scissors are studied by means of appearance and metallography.The results show that bad low-temperature flowing capability of steel and no tempering at high-temperature after forging cause the forging defects.Flowing capability of steel should be increased by means of increasing forging temperature to avoiding mould cavity non-saturate,crack and pucker.Shearing edge avulsion can be removed by means of tempering at high-temperature after forging.Selecting suitable mould material to reduce the forging stress can increase the mould service life.
Funds:
AuthorIntro:
Reference:
[1]《理化检验》编辑部.金相检验[M].上海:机械工业部上海材料研究所,1983.
[2]王笑天.金属材料学[M].北京:机械工业出版社,1987.
Service:
This site has not yet opened Download Service】【Add Favorite
Copyright Forging & Stamping Technology.All rights reserved
 Sponsored by: Beijing Research Institute of Mechanical and Electrical Technology; Society for Technology of Plasticity, CMES
Tel: +86-010-62920652 +86-010-82415085     Fax:+86-010-62920652
Address: No.18 Xueqing Road, Beijing 100083, P. R. China
 E-mail: fst@263.net    dyjsgg@163.com